FDA Brief

Company record

Philips North America

0 FDA warning letters since 2023 and 102 recalls in the last 12 months.

Get an email the day FDA posts anything new about Philips North America.

Recalls

ReportedClassFirmProduct and reason
Nov 4 Class II Philips North America
Device
Model Number: 712010.
Potential for the device to fail immediately after an exposure, causing the system to remain stationary.
Oct 28 Class II Philips North America
Device
Philips Smart-hopping 2.0 AP 1.4 GHz; Product/Model Number: 867216;
Unreliable wireless roaming behavior of Smart-hopping 1.0 wireless clients during transport when connecting to Smart-hopping 2.0 infrastructure, resulting in loss of connectivity to the central monitoring system (PIC iX).
Oct 7 Class II Philips North America
Device
Ingenia 3.0T and 3.0T CX; Model number (REF): 1. 781342 (3.0T), 2. 781377 (3.0T), 3. 782103 (3.0T), 4. 781271 (3.0T CX); Software Versions: R11, R12 and R13;
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.
Oct 7 Class II Philips North America
Device
MR 7700; Model number (REF): 1. 782120, 2. 782153, 3. 782155, 4. 782199; Software Versions: R11, R12 and R13;
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.
Oct 7 Class II Philips North America
Device
Ingenia 1.5T, 1.5T CX, and 1.5T S; Model number (REF): 1. 781315 (1.5T), 2. 781341 (1.5T), 3. 781396 (1.5T), 4. 782115 (1.5T), 5. 782140 (1.5T), 6. 781262 (1.5T CX), 7. 781347 (1.5T S); Software Versions: R11, R12 and…
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.
Oct 7 Class II Philips North America
Device
Evolution upgrade 1.5T amd Evolution upgrade 3.0T; Model number (REF): 1. 782116 (1.5T), 2. 782148 (1.5T), 3. 782166 (1.5T), 4. 782201(1.5T), 5. 782117 (3.0T), 6. 782143 (3.0T), 7. 782162 (3.0T); Software Versions: R1…
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.
Oct 7 Class II Philips North America Llc
Device
Philips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333
The Bolus tracking scan application may crash with a system error message. Bolus Tracker shots may continue to occur even if the contrast threshold is reached. This may result in the patient receiving additional radiation.
Oct 7 Class II Philips North America
Device
Upgrade dStream; Model number (REF): 782127; Software Versions: R11, R12 and R13;
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.
Oct 7 Class II Philips North America
Device
MR 5300; Model number (REF): 1. 782110, 2. 782152, 3. 782156, 4. 782196; Software Versions: R11, R12 and R13;
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.
Oct 7 Class II Philips North America
Device
SmartPath to MR 7700; Model number (REF): 1. 782178, 2. 782207; Software Versions: R11, R12 and R13;
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.
Oct 7 Class II Philips North America
Device
Ingenia Blue Seal SE and Blue Seal XE; Model number (REF): 1. 782185 (SE), 2. 782192 (XE); Software Versions: R11, R12 and R13;
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.
Oct 7 Class II Philips North America
Device
Ingenia Ambition S and Ingenia Ambition X.; Model number (REF): 1. 781359 (S), 2. 782108 (S), 3. 782139 (S), 4. 782159 (S), 5. 782194 (S), 6. 781356 (X), 7. 782109 (X), 8. 782138 (X), 9. 782160 (X), 10. 782195 (X); Sof…
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.
Oct 7 Class II Philips North America
Device
Ingenia Elition S and Ingenia Elition X.; Model number (REF): 1. 781357 (S), 2. 782106 (S), 3. 782137 (S), 4. 782157 (S), 5. 781358 (X), 6. 782107 (X), 7. 782119 (X), 8. 782136 (X), 9. 782158 (X), 10. 782198 (X); Softw…
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.
Oct 7 Class II Philips North America
Device
Upgrade to MR 7700; Model number (REF): 782130; Software Versions: R11, R12 and R13;
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.
Oct 7 Class II Philips North America
Device
SmartPath to dStream for 1.5T, SmartPath to dStream for 3.0T, SmartPath to dStream for XR and 3.0T, and SmartPath to Ingenia Elition X; Model number (REF): 1. 781260 (1.5T), 2. 782112 (1.5T), 3. 782146 (1.5T), 4. 782165…
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.
Sep 30 Class II Philips North America Llc
Device
Achieva 3.0T; Model Numbers (REF): 1. 781177, 2. 781277, 3. 781278, 4. 781344, 5. 781345;
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
Sep 30 Class II Philips North America Llc
Device
Ingenia Elition S and Ingenia Elition X; Model Numbers (REF): 1. 781357(S), 2. 782106(S), 3. 782137(S), 4. 781358(X), 5. 782107(X), 6. 782119(X), 7. 782136(X), 8. 782158(X);
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
Sep 30 Class II Philips North America Llc
Device
MR 7700 and Upgrade to MR 7700; Model Number (REF): 1. 782120, 2. 782153, 3. 782130 (Upgrade);
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
Sep 30 Class II Philips North America Llc
Device
MR 5300 Model Number (REF): 1. 782110, 2. 782152;
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
Sep 30 Class II Philips North America Llc
Device
Ingenia 1.5T, Ingenia 1.5T CX and Ingenia 1.5T S; Model Numbers (REF): 1. 781315(1.5T), 2. 781341(1.5T), 3. 781396(1.5T), 4. 782101(1.5T), 5. 782115(1.5T), 6. 782140(1.5T), 7. 781261(1.5T CX), 8. 781262(1.5T CX), 9. 781…
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
Sep 30 Class II Philips North America Llc
Device
Achieva XR; Model Numbers (REF): 1. 781153, 2. 781253;
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
Sep 30 Class II Philips North America Llc
Device
Evolution upgrade 1.5T and Evolution upgrade 3.0T; Model Numbers (REF): 1. 782116 (1.5T), 2. 782148 (1.5T), 3. 782117 (3.0T), 4. 782143 (3.0T);
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
Sep 30 Class II Philips North America Llc
Device
Ingenia 3.0T and Ingenia 3.0T CX; Model Numbers (REF): 1. 781342(3.0T), 2. 781377(3.0T), 3. 782103(3.0T), 4. 781271(3.0T CX), 5. 782105(3.0T CX);
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
Sep 30 Class II Philips North America Llc
Device
Name (Model Number): 1. Intera 1.5T (781195), 2. Intera 1.5T (781295), 3. Intera 1.5T Achieva IT Nova (781175), 4. Intera 1.5T Achieva Nova (781172), 5. Intera 1.5T Achieva Nova-Dual (781173), 6. Intera 1.5T Omni/Stellar…
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
Sep 30 Class II Philips North America Llc
Device
Ingenia Ambition S and Ingenia Ambition X; Model Numbers (REF): 1. 781359 (S), 2. 782108 (S), 3. 782139 (S), 4. 781356 (X), 5. 782109 (X), 6. 782138 (X);
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
Sep 30 Class II Philips North America Llc
Device
Achieva 1.5T, Achieva 1.5T Conversion and Achieva 1.5T Initial system; Model Numbers (REF): 1. 781196 (Achieva 1.5T), 2. 781343 (Achieva 1.5T), 3. 781283 (Achieva 1.5T Conversion), 4. 781346 (Achieva 1.5 Conversion), 5.…
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
Sep 30 Class II Philips North America Llc
Device
SmartPath to dStream for 1.5T, SmartPath to dStream for 3.0T, SmartPath to dStream for XR and 3.0T, SmartPath to Ingenia Elition X; Model Numbers (REF): 1. 781260 (1.5T), 2. 782112 (1.5T), 3. 782146 (1.5T), 4. 782145 (3…
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
Sep 23 Class II Philips North America
Device
Philips PIC iX Essentials REF:867093
Due to software malfunction, patient information system may freeze and become unresponsive.
Sep 23 Class II Philips North America
Device
Philips Patient Information Center iX REF:866389
Due to software malfunction, patient information system may freeze and become unresponsive.
Sep 9 Class II Philips North America Llc
Device
Philips Brilliance CT 64: Brilliance CT 64 Channel Product Code: 728231, Brilliance CT 64 Upgrades Product Code: 728232. Software 4.1.10.x
Philips has identified an unintended radiation issue with the Philips IQon, iCT, Ingenuity, and Brilliance CT 64 systems that could affect the performance of the equipment.
Sep 9 Class II Philips North America Llc
Device
Philips Ingenuity CT Product Codes: Ingenuity CT Upgrades 728327, Ingenuity CT China 728324, Ingenuity Core 128 Elite Brazil 728325. Software 4.4.10.x
Philips has identified an unintended radiation issue with the Philips IQon, iCT, Ingenuity, and Brilliance CT 64 systems that could affect the performance of the equipment.
Sep 9 Class II Philips North America Llc
Device
Philips IQon Spectral CT Product Codes: IQon Spectral CT 728332, IQon Spectral CT 728331 Software 4.7.7.x
Philips has identified an unintended radiation issue with the Philips IQon, iCT, Ingenuity, and Brilliance CT 64 systems that could affect the performance of the equipment.
Sep 9 Class II Philips North America Llc
Device
Philips Brilliance. Models: Brilliance iCT, Brilliance iCT Upgrades, Brilliance iCT SP. Software 4.1.10.x Product Codes: 728306, 728305, 728311.
Philips has identified an unintended radiation issue with the Philips IQon, iCT, Ingenuity, and Brilliance CT 64 systems that could affect the performance of the equipment.
Aug 26 Class II Philips North America Llc
Device
Cardiac Workstation 5000. Product Number: 860439.
Device will intermittently provide inaccurate data related to native heart rate.
Aug 26 Class II Philips North America Llc
Device
Cardiac Workstation 7000. Product Number: 860441.
Device will intermittently provide inaccurate data related to native heart rate.
Jul 1 Class II Philips North America Llc
Device
Philips Avalon Fetal Monitor, FM20, Part number M2702A, Part M2703A
Monitor has incorrect assembly of the speaker connector which can cause device cables to be pulled out of the housing and increase the risk of intermittent or permanent loss of speaker output.
Jul 1 Class II Philips North America Llc
Device
Philips Avalon Fetal Monitor, FM 30 Part numberM2703A
Monitor has incorrect assembly of the speaker connector which can cause device cables to be pulled out of the housing and increase the risk of intermittent or permanent loss of speaker output.
Jul 1 Class II Philips North America Llc
Device
Philips Telemetry Monitor 5500 1.4 GHz. Model Number: 867232. Includes the following system descriptions: TELE Monitor 5500, 1.4 Ghz, ECG only, ex; Telemetry Monitor 5500 1.4 GHz FAST EX.
Potential for device reset to default "NEW_DEVICE" state, which may lead to a to loss of device configuration and the equipment label, resulting in a failure to reconnect to the PIC iX central monitoring system.
Jun 17 Class II Philips North America Llc
Device
DigitalDiagnost C90 Flex/Value/Chest/ER. Model Number: 712035.
System was delivered without the required Seismic kit for Wall Stand VS2, which is designed to prevent damage and ensure stability of the Wall Stand VS2 during seismic events.
May 6 Class II Philips North America
Device
Philips SmartPath to dStream for 3.0T with MR Elastography (MRE). 1. Model Number (REF): 782145.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
May 6 Class II Philips North America
Device
Philips SmartPath to Ingenia Elition X with MR Elastography (MRE). 1. Model Number (REF): 782118. 2. Model Number (REF): 782144. 3. Model Number (REF): 782163.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
May 6 Class II Philips North America
Device
Philips Ingenia 3.0T with MR Elastography (MRE). 1. Model Number (REF): 781342. 2. Model Number (REF): 781377. 3. Model Number (REF): 782103. 510(k) Numbers: K163116, K173079, K183063, K193215, K213516, K…
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
May 6 Class II Philips North America
Device
Philips Evolution upgrade 3.0T with MR Elastography (MRE). 1. Model Number (REF): 782143. 2. Model Number (REF): 782162.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
May 6 Class II Philips North America
Device
Philips Ingenia Ambition S with MR Elastography (MRE). 1. Model Number (REF): 781359. 2. Model Number (REF): 782108. 3. Model Number (REF): 782139.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
May 6 Class II Philips North America
Device
Philips Achieva 1.5T with MR Elastography (MRE). 1. Model Number (REF): 781196. 2. Model Number (REF): 781296.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
May 6 Class II Philips North America
Device
Philips Ingenia 1.5T CX with MR Elastography (MRE). 1. Model Number (REF): 781262.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
May 6 Class II Philips North America
Device
Philips Ingenia Elition X with MR Elastography (MRE). 1. Model Number (REF): 781358. 2. Model Number (REF): 782107. 3. Model Number (REF): 782136. 510(k) Numbers: K213516, K223458, K230972, K232030, K…
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
May 6 Class II Philips North America
Device
Philips Ingenia 3.0T CX with MR Elastography (MRE). 1. Model Number (REF): 781271. 2. Model Number (REF): 782105. 510(k) Numbers: K173079, K183063, K193215, K213583, K223458, K251397, K251808.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
May 6 Class II Philips North America
Device
Philips Achieva 1.5T Initial system with MR Elastography (MRE). 1. Model Number (REF): 781178.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
May 6 Class II Philips North America
Device
Philips Ingenia 1.5T with MR Elastography (MRE). 1. Model Number (REF): 781315. 2. Model Number (REF): 781341. 3. Model Number (REF): 781396. 4. Model Number (REF): 782115. 510(k) Numbers: K163116, K1…
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
May 6 Class II Philips North America
Device
Philips MR 7700 with MR Elastography (MRE). 1. Model Number (REF): 782120. 2. Model Number (REF): 782153.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
May 6 Class II Philips North America
Device
Philips Achieva 3.0T with MR Elastography (MRE). 1. Model Number (REF): 781278.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
May 6 Class II Philips North America
Device
Philips Evolution Upgrade 1.5T with MR Elastography (MRE). 1. Model Number (REF): 782116 2. Model Number (REF): 782148. 3. Model Number (REF): 782166.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
May 6 Class II Philips North America
Device
Philips Ingenia Elition S with MR Elastography (MRE). 1. Model Number (REF): 781357. 2. Model Number (REF): 782106. 3. Model Number (REF): 782137.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
May 6 Class II Philips North America
Device
Philips Upgrade to MR 7700 with MR Elastography (MRE). 1. Model Number (REF): 782130.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
May 6 Class II Philips North America
Device
Philips Ingenia Ambition X with MR Elastography (MRE). 1. Model Number (REF): 781356. 2. Model Number (REF): 782109. 3. Model Number (REF): 782138. 4. Model Number (REF): 782160.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
May 6 Class II Philips North America
Device
Philips SmartPath to dStream for XR and 3.0T with MR Elastography (MRE). 1. Model Number (REF): 781270. 2. Model Number (REF): 782113.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
May 6 Class II Philips North America
Device
Philips SmartPath to dStream for 1.5T with MR Elastography (MRE). 1. Model Number (REF): 781260. 2. Model Number (REF): 782112.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
Feb 18 Class II Philips North America Llc
Device
Brilliance iCT; Product Code (REF): 728306;
As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.
Feb 18 Class II Philips North America Llc
Device
Spectral CT; Product Code (REF): 728333;
As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.
Feb 18 Class II Philips North America Llc
Device
IQon Spectral CT; Product Code (REF): 728332;
As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.
Feb 18 Class II Philips North America Llc
Device
Spectral CT 7500 on Rails; Product Code (REF): 728334;
As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.
Feb 4 Class II Philips North America
Device
SmartPath to dStream for 3.0T. Product Code (REF): 782145. MR systems with SW version R11.1 and R12.1.
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
Feb 4 Class II Philips North America
Device
Ingenia Ambition S. Product Code (REF): 782108. MR systems with SW version R11.1 and R12.1.
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
Feb 4 Class II Philips North America
Device
Evolution Upgrade 1.5T. Product Codes (REF): (1) 782148, (2) 782116. MR systems with SW version R11.1 and R12.1.
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
Feb 4 Class II Philips North America
Device
Ingenia 3.0T. Product Code (REF): (1) 781342, (2) 781377. MR systems with SW version R11.1 and R12.1.
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
Feb 4 Class II Philips North America
Device
MR 7700. Product Code (REF): (1) 782120, (2) 782153. MR systems with SW version R11.1 and R12.1.
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
Feb 4 Class II Philips North America
Device
Ingenia 3.0T CX. Product Code (REF): 781271. MR systems with SW version R11.1 and R12.1.
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
Feb 4 Class II Philips North America
Device
Ingenia Elition X. Product Code (REF): (1) 781358, (2) 782107, (3) 782136. MR systems with SW version R11.1 and R12.1.
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
Feb 4 Class II Philips North America
Device
Evolution Upgrade 3.0T. Product Code (REF): 782143. MR systems with SW version R11.1 and R12.1.
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
Feb 4 Class II Philips North America
Device
SmartPath to dStream for XR and 3.0T. Product Code (REF): 781270. MR systems with SW version R11.1 and R12.1.
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
Feb 4 Class II Philips North America
Device
Ingenia Ambition X. Product Code (REF): (1) 781356, (2) 782109, (3)782138. MR systems with SW version R11.1 and R12.1.
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
Feb 4 Class II Philips North America
Device
Ingenia Elition S. Product Code (REF): (1) 781357, (2) 782106, (3)782137. MR systems with SW version R11.1 and R12.1.
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
Feb 4 Class II Philips North America
Device
Upgrade to MR 7700. Product Code (REF): 782130. MR systems with SW version R11.1 and R12.1.
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
Feb 4 Class II Philips North America
Device
Ingenia 1.5T. Product Code (REF): (1) 781341, (2) 781396. MR systems with SW version R11.1 and R12.1.
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
Jan 28 Class II Philips North America
Device
Patient Information Center iX; Software Version Number: 4.5.0
A recent software (SW) patch modifies Mobile Event Notification filter settings without providing any indication to the user when upgrading the system.
Jan 14 Class II Philips North America Llc
Device
Philips Smart-hopping 2.0 AP 1.4 GHz. Patient Monitor.
It was found that the MX40 device could not reconnect to the PIC iX when moving between Radiohead and Trident 1.4 GHz access points if the signal strength changed quickly.
Dec 17, 2025 Class II Philips North America
Device
Azurion R3.0. Model Number 722234. Commercial Name: Azurion 7 M20. INTERVENTIONAL FLUOROSCOPIC XRAY SYSTEM
Between July 2024 and June 2025, Philips Azurion R3.0 systems with a FlexArm stand were shipped without the Source-to-Skin Distance (SSD) Spacer. The SSD spacer is necessary to meet the FDA minimum source-skin distance requirement of 38 cm for standard fluoroscopy applications.
Dec 10, 2025 Class II Philips North America Llc
Device
IntelliVue MP90. Product Number: M8010A.
Potential issue where the IntelliVue monitors did not alarm.
Dec 10, 2025 Class II Philips North America Llc
Device
IntelliVue MP30. Product Number: M8002A.
Potential issue where the IntelliVue monitors did not alarm.
Dec 10, 2025 Class II Philips North America Llc
Device
IntelliVue Patient Monitor MX800. Product Number: 865240.
Potential issue where the IntelliVue monitors did not alarm.
Dec 10, 2025 Class II Philips North America Llc
Device
IntelliVue Patient Monitor MX700. Product Number: 865241.
Potential issue where the IntelliVue monitors did not alarm.
Dec 10, 2025 Class II Philips North America Llc
Device
CT 5300; Software Version Number: 4.5, 5.0, 5.1;
If the screws of tube heat exchanger on the rotating scanner are not tightened well after replacing. This component may become detached and make contact with other components located within the Incisive CT systems during rotation. Other components could be damaged due to the contact with the tube heat exchanger.
Dec 10, 2025 Class II Philips North America Llc
Device
IntelliVue Patient Monitor MX500. Product Number: 866064.
Potential issue where the IntelliVue monitors did not alarm.
Dec 10, 2025 Class II Philips North America Llc
Device
IntelliVue Patient Monitor MX850. Product Number: 866470.
Potential issue where the IntelliVue monitors did not alarm.
Dec 10, 2025 Class II Philips North America Llc
Device
IntelliVue MP2. Product Number: M8102A.
Potential issue where the IntelliVue monitors did not alarm.
Dec 10, 2025 Class II Philips North America Llc
Device
Incisive CT; Software Version Number: 4.5, 5.0, 5.1;
If the screws of tube heat exchanger on the rotating scanner are not tightened well after replacing. This component may become detached and make contact with other components located within the Incisive CT systems during rotation. Other components could be damaged due to the contact with the tube heat exchanger.
Dec 10, 2025 Class II Philips North America Llc
Device
IntelliVue MP50. Product Number: M8004A.
Potential issue where the IntelliVue monitors did not alarm.
Dec 10, 2025 Class II Philips North America Llc
Device
IntelliVue MP80. Product Number: M8008A.
Potential issue where the IntelliVue monitors did not alarm.
Dec 10, 2025 Class II Philips North America Llc
Device
IntelliVue Patient Monitor MX100. Product Number: 867033.
Potential issue where the IntelliVue monitors did not alarm.
Dec 10, 2025 Class II Philips North America Llc
Device
IntelliVue MP40. Product Number: M8003A.
Potential issue where the IntelliVue monitors did not alarm.
Dec 10, 2025 Class II Philips North America Llc
Device
IntelliVue Multi Measurement Server X2. Product Number: M3002A.
Potential issue where the IntelliVue monitors did not alarm.
Dec 10, 2025 Class II Philips North America Llc
Device
IntelliVue MP60. Product Number: M8005A.
Potential issue where the IntelliVue monitors did not alarm.
Dec 10, 2025 Class II Philips North America Llc
Device
IntelliVue MP5. Product Number: M8105A.
Potential issue where the IntelliVue monitors did not alarm.
Dec 10, 2025 Class II Philips North America Llc
Device
IntelliVue MP20. Product Number: M8001A.
Potential issue where the IntelliVue monitors did not alarm.
Dec 10, 2025 Class II Philips North America Llc
Device
IntelliVue Patient Monitor MX600. Product Number: 865242.
Potential issue where the IntelliVue monitors did not alarm.
Dec 10, 2025 Class II Philips North America Llc
Device
IntelliVue Patient Monitor MX400. Product Number: 866060.
Potential issue where the IntelliVue monitors did not alarm.
Dec 10, 2025 Class II Philips North America Llc
Device
IntelliVue Multi-Measurement Module X3. Product Number: 867030.
Potential issue where the IntelliVue monitors did not alarm.
Dec 10, 2025 Class II Philips North America Llc
Device
IntelliVue Patient Monitor MX550. Product Number: 866066.
Potential issue where the IntelliVue monitors did not alarm.
Dec 10, 2025 Class II Philips North America Llc
Device
IntelliVue Patient Monitor MX750. Product Number: 866471.
Potential issue where the IntelliVue monitors did not alarm.
Dec 10, 2025 Class II Philips North America Llc
Device
IntelliVue Patient Monitor MX450. Product Number: 866062.
Potential issue where the IntelliVue monitors did not alarm.
Dec 10, 2025 Class II Philips North America Llc
Device
IntelliVue MP70. Product Number: M8007A.
Potential issue where the IntelliVue monitors did not alarm.